Fluoro-saver marker
Distal tip 15mm
Distal delivery system
30% surface metal coverage
Cobalt Chromium, 64 wires with 12 radiopaque wires
WORKING VIEWS
Check your working views
SIZING AND PLACEMENT
Keep lenght > 5mm on either side of the neck
PREPARE YOUR DEVICE
Insert to mid-valve
Flush with 3 or 5ml syringe
Introduce sheath to the bottom of the valve
Push the device into the microcatheter
Until fluoro-saver marker
DEPLOYMENT
Low push forces through microcatheter even in the tortuous anatomy
Check the position of the distal tip
To resheath the implant : advance the microcatheter and gently pull back delivery wire
If positioning is not satisfactory, you can recapture the implant up to 3 times
Maintain the delivery wire within the center of the parent vessel during deployment
Remove any build-up of slack in the system